Testomat® LAB TH

Product number: 116102

€3,590.80*

Please log in to receive your conditions.

Available, delivery time: 1-2 days

Version

Product number: 116102
GTIN: 4260682383778
Measurement parameters: Water hardness (residual total hardness)
Measuring range: 0.89 - 448 ppm
Menu navigation: (Servicemonitor) German, English, French, Dutch, Portuguese
Would you like to register the warranty for this product?
Register warranty here
A warranty extension is possible after taking out an indicator subscription.
Offer as PDF  
Login for addressed offer
The entered number of pieces will be taken into account when creating the offer.
Product information "Testomat® LAB TH"
The Testomat® LAB TH and Testomat® LAB TH-R (Remote)determine the water hardness (residual total hardness) fully automatically by titration. The instrument is suitable for controlling the water quality of water treatment plants, drinking water plants, industrial boilers and for monitoring process water.

Please note that Testomat® LAB - TH and Testomat® LAB TH-R was developed for use in multi-parameter systems or for connection to a higher-level control system.
In addition to the 4-20 mA output for transmitting the measured values, this series has an RS232 interface for transmitting the measured values and error & status messages. Furthermore, all data are continuously recorded on the integrated SD card or optionally available SDHC card (2Gbyte) as a file in CSV format structured and can be used at any time for detailed observations.

The Testomat® LAB TH and TH-R convinces with new functionalities:
  • Execution of parameterization by SD card or the "Service Monitor" software via mini-USB (on the device side) to USB 2.0 (for example PC/notebook)
  • Multilingual menu navigation software in German, English, French, Dutch, Portuguese
  • Firmware update via SD card
  • compact design
  • optimized water detection based on an optical measuring method
  • Integrated self-test with continuous monitoring
  • Automatic venting of the indicator line

Additional functions of the TH-R version:
  • Parameterisation and readout of the unit parameters via the RS232 interface

Tried and tested performance profile:
  • High-precision titration using a piston dosing pump
  • Reliable and low-maintenance operation
  • Minimal indicator and water consumption
  • Analog output 4 - 20 mA for transmission of measured values
  • Programmable hardness unit in °dH, °f, ppm CaCO3, mmol/l
  • Common alarm output
  • Indicator quantities: 500 ml bottles

Analysis triggering:
  • Automatic interval operation (interval pause adjustable from 0-255 minutes)
  • External analysis input (start/stop)
  • Hand launch

Optional accessories:
  • OLED display
  • Protecting cover to increase the IP protection class

Dimensions

Height 21 cm
Length 43 cm
Width 35 cm

Weight

Weight 4,55 kg

Other attributes

Country of origin EU
Customs tariff number 90278990
Technical specifications
Operating voltage: 24 VDC
Power consumption: Max. 1 A, without external load
Protection class: I
Protection: IP 40 (when using the optionally available device cover, IP43)
Ambient temperature: 10–40 °C
Water temperature: 10–40 °C
Dimensions (W x H x D):     270 x 350 x 147 mm
Operating pressure: 1 to 8 bar / 1x105 to 8x105 Pa or 0.3* to 1 bar / 1x105 to 1x105 Pa (*after removing the regulator core)
Weight: approx. 4.35 kg
Menu Language: (Servicemonitor) German, English, French, Dutch
Measuring range:0.89 - 448 ppm